Evaluation of point‐of‐care capillary and venous blood glucose concentrations in hospitalized neonatal foals

Author:

Dias Moreira Ana Sofia1ORCID,Weng Hsin‐Yi2,Hostnik Laura D.3,Beasley Erin M.4,Peek Simon F.5,Munsterman Amelia S.1ORCID

Affiliation:

1. Department of Large Animal Clinical Sciences Michigan State University East Lansing Michigan USA

2. Department of Comparative Pathobiology Purdue University West Lafayette Indiana USA

3. Department of Veterinary Clinical Sciences The Ohio State University Columbus Ohio USA

4. Department of Large Animal Medicine University of Georgia Athens Georgia USA

5. Department of Medical Sciences University of Wisconsin–Madison Madison Wisconsin USA

Abstract

AbstractObjectiveTo compare glucose measurements from capillary and venous blood samples using a point‐of‐care (POC) glucometer with a standard laboratory (colorimetric, glucose oxidase) assay (LABGLU) in a population of hospitalized, neonatal foals.DesignMulticenter, prospective, experimental study, conducted between March 2019 and June 2020.SettingFour university teaching hospitals and 1 private referral hospital.AnimalsFifty‐four hospitalized neonatal (≤30 days of age) foals.InterventionsSimultaneous capillary (muzzle, POCMUZ) and venous (jugular, POCJUG) blood samples were obtained to determine POC glucose concentrations. Venous samples were also analyzed by LABGLU. Each foal was sampled at the time of enrollment or admission to the hospital and at 1 subsequent point during hospitalization. Indirect mean arterial pressure and hematocrit were concurrently recorded.Measurements and Main ResultsBland–Altman analysis showed a mean bias (95% limits of agreement) of −28.0 (−88.6 to 32.6) mg/dL for comparison of POCJUG with LABGLU, −8.2 (−94.3 to 78.0) mg/dL for POCMUZ and LABGLU, and 18.8 (−44.4 to 82.0) mg/dL for POCMUZ and POCJUG. A total of 63.5% of the POCJUG and 45.2% of the POCMUZ samples exceeded the reference value by ±15 mg/dL (for LABGLU samples <75 mg/dL) or ±15% (for LABGLU samples ≥75mg/dL). Concordance correlation coefficient (95% confidence interval [CI]) indicated a fair agreement between POCJUG and LABGLU (0.75, 95% CI: 0.66–0.82) and between POCMUZ and LABGLU (0.71, 95% CI: 0.58–0.80). Fifty percent (14/28) of hypoglycemic foals on the reference method were incorrectly classified as euglycemic by POCJUG, and 5 of 28 were incorrectly classified by POCMUZ.ConclusionsIn the sampled population, the chosen POC glucometer lacked agreement with the standard laboratory measurement. Limits of agreement were wide for both POCJUG and POCMUZ. Inaccuracies in POC results could impact decision‐making in the clinical management of glycemic control in hospitalized neonatal foals and, importantly, increase the risk of hypoglycemic events being underdiagnosed in critical patients.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.7亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2025 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3