A prospective comparison of invasive and non‐invasive blood pressure in children undergoing cardiac catheterization

Author:

Beringer Richard1,Keith Alastair1,Jones Elin2,Murphy Tim1ORCID,White Paul3ORCID

Affiliation:

1. Bristol Royal Hospital for Children Bristol UK

2. Birmingham Children's Hospital Birmingham UK

3. University of the West of England Bristol UK

Abstract

AbstractBackgroundBlood pressure measurement is a standard of monitoring during general anesthesia. Invasive measurement is considered the gold standard but is less commonly used than non‐invasive. Automated oscillometric blood pressure devices measure the mean arterial pressure (MAP) and use an algorithm to determine the systolic and diastolic pressures. Few devices have been validated in children, particularly during anesthesia. Few studies have assessed the agreement between invasive and non‐invasive blood pressure measurements in children.MethodsThis was a multi‐center prospective observational study of children under 16 years undergoing cardiac catheterization with general anesthesia. Paired invasive and non‐invasive blood pressure measurements were recorded for each patient during stable periods of the procedure. Correlation within and between sites was assessed with Pearson's correlation coefficient, and agreement was examined using Bland–Altman methodology to determine bias. Agreement during episodes of hypotension and for age and weight was also determined. Bias greater than 5 mmHg and standard deviation greater than 8 mmHg was considered clinically significant. The primary end point was agreement of MAP measurements.ResultsA total of 683 paired blood pressure values were collected from 254 children in three pediatric hospitals. Median [IQR] age and weight were 3 [1–7] years and 13.9 [8–23] Kg. The overall bias (SD) for mean arterial pressure values was 7.2 (11.4) mmHg. During hypotension (190 readings), the bias (SD) was 15 (11.0) mmHg. The non‐invasive MAP was frequently higher than invasive MAP during infancy, and lower in older children.ConclusionAutomated oscillometric blood pressure measurement is unreliable in anesthetized children during cardiac catheterization. Invasive pressure measurement should be considered for high‐risk cases.

Publisher

Wiley

Subject

Anesthesiology and Pain Medicine,Pediatrics, Perinatology and Child Health

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3