In vitro hemolytic performance of the Realheart®V11C TAHprototype with porcine blood

Author:

Perkins Ina Laura1ORCID,Mansisidor Eva Maria1ORCID,Zaman Shaikh Faisal1ORCID,Mudge Kyle1ORCID,Bodger Owen2ORCID,Najar Azad1ORCID

Affiliation:

1. R&D Scandinavian Real Heart AB Västerås Sweden

2. Institute of Life Science Swansea University Medical School Swansea UK

Abstract

AbstractBackgroundHemolysis testing of new devices to treat heart failure is a regulatory requirement. The ASTM F1841–97 standard for hemolysis testing was developed for continuous flow pumps and does not specify test rig design. When research groups use different methodologies, results are difficult to compare. Pulsatile flow pump rigs require compliance chambers, and thus, the Aachen rig (Gräf et al) was developed for the pulsatile Reinheart TAH. The study objective was to use this rig to test the early Realheart TAH prototype V11C hemolysis performance compared to literature.MethodsThe experimental control was the continuous flow pump BPX‐80 (Medtronic) and pooled heparinized porcine blood was used.ResultsThe mgNIH of BPX‐80 and V11C was 5.42 ± 1.47 and 25.20 ± 5.46 mg/100 L, respectively. The NIH ratio of V11C over BPX‐80 was 5.5.ConclusionThe absolute and the relative hemolysis of the V11C are lower compared to both the large and small Reinheart TAH devices published values. Pulsatile pumps create more hemolysis in the Aachen rig, and it is not known if this is because how the rig handles pulsatile flow or due to the devices. Future studies will, therefore, use a pulsatile pump such as the SynCardia as clinical comparator and human blood to test the performance of future Realheart TAH prototypes.

Funder

Scandinavian Real Heart

Publisher

Wiley

Subject

Biomedical Engineering,General Medicine,Biomaterials,Medicine (miscellaneous),Bioengineering

Reference29 articles.

1. Global epidemiology and future trends of heart failure

2. Trends in survival after a diagnosis of heart failure in the United Kingdom 2000-2017: population based cohort study

3. Heart Disease and Stroke Statistics—2021 Update

4. Indexes of hemolysis in human recipients of the Jarvik‐7 total artificial heart: a cooperative report of fifteen patients;Levinson MM;J Heart Transplant,1986

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3