Tissue Temperatures and Lesion Size During Irrigated Tip Catheter Radiofrequency Ablation: An In Vitro Comparison of Temperature‐Controlled Irrigated Tip Ablation, Power‐Controlled Irrigated Tip Ablation, and Standard Temperature‐Controlled Ablation

Author:

PETERSEN HELEN HØGH,CHEN XU,PIETERSEN ADRIAN,SVENDSEN JESPER HASTRUP,HAUNSØ STIG

Abstract

The limited success rate of radiofrequency catheter ablation in patients with ventricular tachycardias related to structural heart disease may be increased by enlarging the lesion size. Irrigated tip catheter ablation is a new method for enlarging the size of the lesion. It was introduced in the power‐controlled mode with high power and high infusion rate, and is associated with an increased risk of crater formation, which is related to high tissue temperatures. The present study explored the tissue temperatures during temperature‐controlled irrigated tip ablation, comparing it with standard temperature‐controlled ablation and power‐controlled irrigated tip ablation. In vitro strips of porcine left ventricular myocardium were ablated. Temperature‐controlled irrigated tip ablation at target temperatures 60°C, 70°C, and 80°C with infusion of 1 mL saline/min were compared with standard temperature‐controlled ablation at 70°C and power‐controlled irrigated tip ablation at 40 W, and infusion of 20 mL/min. Lesion size and tissue temperatures were significantly higher during all modes of irrigated tip ablation compared with standard temperature‐controlled ablation (P < 0.05). Lesion volume correlated positively with tissue temperature (r = 0.87). The maximum recorded tissue temperature was always 1 mm from the ablation electrode and was 67 ± 4°C for standard ablation and 93 ± 6°C, 99 ± 6°C, and 115 ± 13°C for temperature‐controlled irrigated tip ablation at 60°C, 70°C, and 80°C, respectively, and 112 ± 12°C for power‐controlled irrigated tip ablation, which for irrigated tip ablation was significantly higher than tip temperature (P < 0.0001). Crater formation only occurred at tissue temperatures > 100°C. We conclude that irrigated tip catheter ablation increases lesion size and tissue temperatures compared with standard ablation in the temperature‐controlled mode at the same or higher target temperatures and in the power‐controlled mode. Furthermore, tissue temperature and delivered power are the best indicators of lesion volume during temperature‐controlled ablation.

Publisher

Wiley

Reference19 articles.

1. Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle Preparation

2. Myocardial healing patterns during radiofrequency energy delivery via an irrigated electrode;Demazumder DBS;PACE,1997

3. Comparison of different electrodes for radiofrequency ablation of myocardium using saline “irrigation”;Demazumdfir DBS;PACE,1997

4. Impact of ablation location on impedance pop occurrence with irrigated tip energy delivery in dogs;Packer DL;PACE,1997

5. Radiofrequency catheter ablation of ventricular myocardium using a cooled catheter system;Dorwarth U;Eur Heart f,1996

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.7亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2025 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3