Evaluation of a retinal oximetry image quality indicator in patients with cataract

Author:

Klefter Oliver Niels12ORCID,Erichsen Jesper Høiberg1,Hansen Mathias Møller1ORCID,Holm Lars Morten12,Hardarson Sveinn Hakon3,Stefánsson Einar34ORCID,Kessel Line12ORCID

Affiliation:

1. Department of Ophthalmology Rigshospitalet Glostrup Copenhagen Denmark

2. Department of Clinical Medicine University of Copenhagen Copenhagen Denmark

3. Department of Medicine University of Iceland Reykjavik Iceland

4. Department of Ophthalmology Landspitali Reykjavik Iceland

Abstract

AbstractPurposeTo evaluate a new automated retinal oximetry image quality indicator with cataract as a clinical model.MethodsSixty‐one eyes in 61 patients were imaged by the Oxymap T1 Retinal Oximeter at baseline and 25 eyes were also examined 3 weeks after cataract surgery. Image quality (0–10 on a continuous scale) was compared with standardized AREDS cataract grading and Pentacam lens densitometry. Associations with retinal oximetry measurements and visual acuity were examined.ResultsImage quality correlated with total, nuclear and posterior subcapsular cataract grades (ANOVA, p < 0.05), tended to be associated with lens densitometry and it improved from 4.3 ± 1.4 to 5.7 ± 1.0 (p < 0.05) after cataract surgery. Very low image quality, below 3, led to vessel detection failure in retinal oximetry images. Higher image qualities were linearly associated with higher measured retinal oxygen saturations (r = 0.52 in arteries and r = 0.46 in veins; p < 0.001).ConclusionRetinal oximetry image quality deteriorated with increasing cataract density and improved after cataract surgery, supporting its use as a measure of optical clarity. The numerical quality indicator demonstrated a threshold below which images of poor optical quality should be discarded. Image quality affects the estimates of retinal oximetry parameters and should therefore be included in future analyses.

Publisher

Wiley

Subject

Ophthalmology,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3