Abstract
We compared the accuracy and precision of three devices for measuring soil water content in both natural and repacked soils and evaluated their temperature sensitivity. Calibrations were developed for a capacitance instrument (ECH2O), a time domain reflectometry cable tester (CT), and a water content reflectometer (WCR) in soils collected from the Wind River and H.J. Andrews Experimental Forests. We compared these calibrations with equations suggested by manufacturers or commonly used in the literature and found the standard equations predicted soil moisture content 0%11.5% lower (p < 0.0001) than new calibrations. Each new calibration equation adequately predicted soil moisture from the output for each instrument regardless of location or soil type. Prediction intervals varied, with errors of 4.5%, 3.5%, and 7.1% for the ECH2O, CT, and WCR, respectively. Only the ECH2O system was significantly influenced by temperature for the range sampled: as temperature increased by 1 °C, the soil moisture estimate decreased by 0.1%. Overall, the ECH2O performed nearly as well as the CT, and thanks to its lower cost, small differences in performance might be offset by deployment of a greater number of probes in field sampling. Despite its higher cost, the WCR did not perform as well as the other two systems.