Affiliation:
1. Johns Hopkins University
2. University of Birmingham
3. Dalian University of Technology
Abstract
Due to low imaging contrast, a widely-used cone-beam computed tomography-guided small animal irradiator is less adept at localizing in vivo soft tissue targets. Bioluminescence tomography (BLT), which combines a model of light propagation through tissue with an optimization algorithm, can recover a spatially resolved tomographic volume for an internal bioluminescent source. We built a novel mobile BLT system for a small animal irradiator to localize soft tissue targets for radiation guidance. In this study, we elaborate its configuration and features that are indispensable for accurate image guidance. Phantom and in vivo validations show the BLT system can localize targets with accuracy within 1 mm. With the optimal choice of threshold and margin for target volume, BLT can provide a distinctive opportunity for investigators to perform conformal biology-guided irradiation to malignancy.
Funder
Xstrahl Ltd
Cancer Prevention and Research Institute of Texas
National Cancer Institute
Subject
Atomic and Molecular Physics, and Optics,Biotechnology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献