Affiliation:
1. Central Laser Facility, CCLRC Rutherford Appleton Laboratory, Didcot, Oxfordshire, OX11 0QX, UK
Abstract
We compare and contrast two Raman collection geometries, backscattering and transmission, to identify their potential for monitoring the bulk chemical composition of turbid media. The experiments performed on pharmaceutical tablets confirm the expected strong bias of the backscattering Raman collection towards surface layers of the probed sample. However, this bias is largely absent with the transmission geometry, exhibiting gross insensitivity to the depth of impurities within the sample. The results are supported by Monte-Carlo simulations. The applicability of transmission geometry to tablets without any thinning is possible because of long migration times of Raman photons in non-absorbing powder media. The absolute measured intensity of the Raman signal was only 12 times lower in transmission geometry compared with backscattering geometry for a standard paracetamol tablet with a thickness of 3.9 mm. This makes detection relatively straightforward, and detectable Raman signals were observed even after propagation through three paracetamol tablets. Given its properties and instrumental simplicity, the transmission method is particularly well suited to the on-line analysis of bulk content of tablets in pharmaceutical applications.
Subject
Spectroscopy,Instrumentation
Cited by
141 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献