Impedance Change as an Indicator of Cochlear Implant Failure

Author:

Pisa Justyn F. D.1,Almutairi Walleed H.1,Mackie Kristy2,Stangherlin Daniela2,Sulkers Jacob2,Hochman Jordan B.1

Affiliation:

1. University of Manitoba

2. Central Speech and Hearing Clinic, Winnipeg, Manitoba, Canada

Abstract

Objective In February 2020, Advanced Bionics initiated a field action notice to remove the MS Ultra (V1) implant from circulation. In this study, we quantify a single site’s incidence with device failure and examine the relationship between impedance change and declining speech perception. Study Design Retrospective/cohort study. Setting Tertiary healthcare center. Patients Forty-nine adult patients (52 devices) were implanted between October 2017 and December 2019, with the following exclusion criteria: <18 years of age, medical/surgical failures, and lack of English-language proficiency. Interventions Diagnostic. Main Outcome Measures Postoperative speech perception (AzBio sentence test) scores at 12 months postactivation were compared with repeat testing at 6- to 12-month intervals. Degree of change in impedances from 1 month postactivation was analyzed at similar intervals. Device failures were suspected when impedance levels on three or more electrodes dropped to <3.5 kΩ and AzBio scores declined >15%. Device failures were confirmed through analyses completed by the manufacturer. Results To date, 21% (11/52) of devices have met all three criteria for a confirmed failure with an average decline in speech perception of 27.7 ± 22%. The average length of time for detection of device failure was 21.9 ± 5 months postimplantation. An additional 7 devices (13%) are currently being monitored for failure but have not yet met full criteria. There was no predictive relationship between degree of impedance change and speech perception scores. Electrodes along the entire array (channels 1–16) can be affected. Conclusions Impedance values can be used as a reliable indicator of device malfunction, thus facilitating patient counseling and early intervention. No correlation was identified between impedance changes and speech perception scores.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Neurology (clinical),Sensory Systems,Otorhinolaryngology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Osteoneogenesis at the Round Window: A Possible Cause of Cochlear Implant Failure?;Audiology Research;2023-12-21

2. The Assessment of Electrode-Neuron Interface (ENI) in Cochlear Implant Users;Latest Advances in Cochlear Implant Technologies and Related Clinical Applications;2023-10-04

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司