Efficacy and Safety of an Adjustable Glaucoma Drainage Device (eyeWatch System) for Refractory Glaucoma: A 2-Year Prospective Study

Author:

Roy Sylvain12,Mermoud André2

Affiliation:

1. Laboratory of Hemodynamics and Cardiovascular Technology, Swiss Federal Institute of Technology

2. Glaucoma Center, SwissVisio, Montchoisi Clinic, Lausanne, Switzerland

Abstract

Précis: In this study, we report a 2-year follow-up after implantation of an adjustable glaucoma drainage device combined with a novel orbital filtering plate. The IOP was efficiently controlled postoperatively with limited complications and an excellent safety profile. Purpose: To report the clinical results at 2 years after implantation of an eyeWatch (eW) system, which comprises an adjustable glaucoma drainage device (eW implant) combined with a novel orbital filtering plate (eyePlate), in refractory glaucoma. Patients and Methods: Monocentric, prospective, noncomparative clinical trial. Patients suffering from refractory glaucoma after several failed surgeries were operated using the eW system. The primary outcome was the success rate, defined as an IOP ≤18 mm Hg and reduction of more than 20% from baseline, IOP ≥6 mm Hg. Secondary outcomes were mean IOP, visual acuity, the number of antiglaucoma medications, and the number and type of complications. Results: Forty-two eyes from 42 patients were included. The mean follow-up time was 22.0±4.0 months. The mean baseline IOP decreased from 23.2±7.2 mm Hg before surgery to 11.9±4.4 mm Hg at 12 months (P<0.001) and 11.5±3.0 at 24 months. The mean number of glaucoma medications decreased from 2.9±0.9 before surgery to 0.7±0.9 at 12 months (P<0.001) and 1.0±0.9 at 24 months. The qualified success rate was 93% at 12 months and 90% at 24 months whereas the complete success rate was 46% and 40% at 12 and 24 months, respectively. Complications occurred in 7 patients. Conclusions: Implantation of the eW system effectively lowered IOP and reduced glaucoma medications for a 2-year period in patients who had previously failed glaucoma surgery. Persistent hypotony was not observed and the number of complications was low.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Ophthalmology

Reference19 articles.

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. A new self-adjustable glaucoma valve;Frontiers in Bioengineering and Biotechnology;2024-05-02

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3