Multicomponent Mathematical Model for Tumor Volume Calculation with Setup Error Using Single-Isocenter Stereotactic Radiotherapy for Multiple Brain Metastases

Author:

Nakano Hisashi1ORCID,Shiinoki Takehiro2,Tanabe Satoshi3,Nakano Toshimichi3,Takizawa Takeshi3,Utsunomiya Satoru3,Sakai Madoka3,Tanabe Shunpei3,Ohta Atsushi3,Kaidu Motoki3,Nishio Teiji4,Ishikawa Hiroyuki3

Affiliation:

1. Niigata University

2. Yamaguchi University: Yamaguchi Daigaku

3. Niigata University: Niigata Daigaku

4. Osaka University: Osaka Daigaku

Abstract

Abstract Purpose We evaluated the tumor residual volumes considering six degrees-of-freedom (6DoF) patient setup errors in stereotactic radiotherapy (SRT) with multicomponent mathematical model using single-isocenter irradiation for brain metastases. Methods Simulated spherical gross tumor volumes (GTVs) with 1.0 (GTV 1), 2.0 (GTV 2), and 3.0 (GTV 3)-cm diameters were used. The distance between the GTV center and isocenter (d) was set at 0–10 cm. The GTV was simultaneously translated within 0–1.0 mm (T) and rotated within 0°–1.0° (R) in the three axis directions using affine transformation. We optimized the tumor growth model parameters using measurements of non-small cell lung cancer cell lines' (A549 and NCI-H460) growth. We calculated the GTV residual volume at the irradiation's end using the physical dose to the GTV when the GTV size, d, and 6DoF setup error varied. The d-values that satisfy tolerance values (10%, 35%, and 50%) of the GTV residual volume rate based on the pre-irradiation GTV volume were determined. Results The larger the tolerance value set for both cell lines, the longer the distance to satisfy the tolerance value. Conclusions In GTV residual volume evaluations based on the multicomponent mathematical model on SRT with single-isocenter irradiation, the smaller the GTV size and the larger the distance and 6DoF setup error, the shorter the distance that satisfies the tolerance value might need to be.

Publisher

Research Square Platform LLC

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3