Evaluation of two benchtop blood gas analyzers for measurement of electrolyte concentrations in venous blood samples from dogs

Author:

Schmitz Katrina L.1,Jeffery Unity2,Heinz Justin A.1,Rutter Christine R.1

Affiliation:

1. From the Department of Small Animal Clinical Sciences, College of Veterinary Medicine and Biological Sciences, Texas A&M University, College Station, TX 77843

2. From the Department of Veterinary Patho-biology, College of Veterinary Medicine and Biological Sciences, Texas A&M University, College Station, TX 77843

Abstract

Abstract OBJECTIVE To assess agreement between 2 benchtop blood gas analyzers developed by 1 manufacturer (BGA 1 and BGA 2 [a newer model with reduced maintenance requirements]) and a reference chemistry analyzer for measurement of electrolyte (sodium, chloride, and potassium) in blood samples from dogs. ANIMALS 17 healthy staff- and student-owned dogs and 23 client-owned dogs admitted to an emergency and intensive care service. PROCEDURES Blood collected by venipuncture was placed in lithium heparin–containing tubes. Aliquots were analyzed immediately with each BGA. Samples were centrifuged, and plasma was analyzed with the reference analyzer. Results for each BGA were compared with results for the reference analyzer by Passing-Bablok regression analysis. Percentage differences between BGA and reference analyzer results were compared with published guidelines for total allowable error. RESULTS Proportional bias was detected for measurement of chloride concentration (slope, 0.7; 95% CI, 0.7 to 0.8), and constant positive bias was detected for measurement of chloride (y-intercept, 34, mmol/L; 95% CI, 16.9 to 38 mmol/L) and potassium (y-intercept, 0.1 mmol/L; 95% CI, 0.1 to 0.2 mmol/L) concentrations with BGA 1. There was no significant bias for measurement of potassium or chloride concentration with BGA 2 or sodium concentration with either BGA. Differences from the reference analyzer result exceeded total allowable error guidelines for ≥ 1 sample/analyte/BGA, but median observed measurement differences between each BGA and the reference analyzer did not. CONCLUSIONS AND CLINICAL RELEVANCE Good agreement with reference analyzer results was found for measurement of the selected electrolyte concentrations in canine blood samples with each BGA.

Publisher

American Veterinary Medical Association (AVMA)

Reference18 articles.

1. Diabetic emergencies in small animals;O'Brien MA,2010

2. Therapeutic approach to electrolyte emergencies;Schaer M,2008

3. Blood gas analyzers;Gonzalez AL,2016

4. ASVCP guidelines: quality assurance for point-of-care testing in veterinary medicine;Flatland B,2013

5. Stat Profile Prime Plus VET Critical Care Analyzer,2019

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.7亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2025 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3