An effcient radiation and position monitoring system is useful for preliminary security inspection and radiation emergency preparedness and response. A high sensitivity and effective monitoring system is needed for these applications. In this study, two position-sensitive prototypes were proposed by using of 8 Multi Pixel Photon Counters (MPPC) and coupled to the cesium iodide with Thallium (CsI(Tl)) scintillator. This inorganic scintillator was chosen for its output wavelength compatibility with MPPC, slightly hygroscopic, high photon transmission, and input energy. For both prototypes, eight MPPCs with di_erent con_gurations were used to obtain active areas of 2.0 × 2.4 cm2 and 2.5 × 4.5cm2, respectively. The prototype has been tested with radioactive (Tl-204 and Co-60). The intensity received by each MPPC’s readout will be recorded using Extended Analogue SiPM Integrated ReadOut Chip (EASIROC) module. Our test showed that both prototypes were sensitive to both beta and gamma sources and were able to reliably determine the sources’ positions. It was shown that the first prototype had greater resolution and performance than the second prototype. This studies may help to improve radiation detection system which is suitable for homeland security, medical field and scanning system.