Author:
Safitri Meilia,Pranaditya Afif,Handoko Brama,Anggoro Sotya
Abstract
Abstract
Sterilization data history is needed to maintain and examine an autoclave performance to minimize the risk of sterilization failure. This study aimed to develop a sterilization data recording system for an autoclave with external data storage using an SD card. The TFT LCD is used as the user interface to make it easier for users to monitor temperature and pressure during the sterilization process. The MAX 31865 temperature sensor and the MPX 5700 AP pressure sensor are used to monitor the temperature and pressure in real-time during the sterilization process. The entire sterilization process of the developed system is controlled using The ATMega 2560 microcontroller. Three testings have been carried out to determine the designed temperature, and the pressure data recording system has been a success, Temperature data recording, Pressure data recording, and Process data recording. Based on the conducting test results, the designed recording system gives an excellent performance to record the temperature, pressure level, and data sterilization process with a success rate of 100%.
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Baby scales with data transfer and database storage using Android;PROCEEDINGS OF THE 9TH INTERNATIONAL SYMPOSIUM ON INNOVATIVE BIOPRODUCTION INDONESIA ON BIOTECHNOLOGY AND BIOENGINEERING 2022: Strengthening Bioeconomy through Applied Biotechnology, Bioengineering, and Biodiversity;2023
2. Simple blood bag storage using peltier;THE 2ND INTERNATIONAL CONFERENCE OF SCIENCE AND INFORMATION TECHNOLOGY IN SMART ADMINISTRATION (ICSINTESA 2021);2022