Seven general radiography x-ray detectors with pixel sizes ranging from 175 to 76 μm: technical evaluation with the focus on orthopaedic imaging

Author:

Marshall N WORCID,Vandenbroucke D,Cockmartin L,Wanninger F,Smet M,Feng Y,Ni Y,Bosmans H

Abstract

Abstract Aim. Flat panel detectors with small pixel sizes general can potentially improve imaging performance in radiography applications requiring fine detail resolution. This study evaluated the imaging performance of seven detectors, covering a wide range of pixel sizes, in the frame of orthopaedic applications. Material and methods. Pixel sizes ranged from 175 (detector A175) to 76 μm (detector G76). Modulation transfer function (MTF) and detective quantum efficiency (DQE) were measured using International Electrotechnical Commission (IEC) RQA3 beam quality. Threshold contrast (C T) and a detectability index (d′) were measured at three air kerma/image levels. Rabbit shoulder images acquired at 60 kV, over five air kerma levels, were evaluated in a visual grading study for anatomical sharpness, image noise and overall diagnostic image quality by four radiologists. The detectors were compared to detector E124. Results. The 10% point of the MTF ranged from 3.21 to 4.80 mm−1, in going from detector A175 to detector G76. DQE(0.5 mm−1) measured at 2.38 μGy/image was 0.50 ± 0.05 for six detectors, but was higher for F100 at 0.62. High frequency DQE was superior for the smaller pixel detectors, however C T for 0.25 mm discs correlated best with DQE(0.5 mm−1). Correlation between C T and the detectability model was good (R 2 = 0.964). C T for 0.25 mm diameter discs was significantly higher for D150 and F100 compared to E124. The visual grading data revealed higher image quality ratings for detectors D125 and F100 compared to E124. An increase in air kerma was associated with improved perceived sharpness and overall quality score, independent of detector. Detectors B150, D125, F100 and G76, performed well in specific tests, however only F100 consistently outperformed the reference detector. Conclusion. Pixel size alone was not a reliable predictor of small detail detectability or even perceived sharpness in a visual grading analysis study.

Funder

Agfa HealthCare

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3