Abstract
Abstract
This paper describes the steps taken to improve the measurement speed of a combined low-coherence and confocal refractive index measurement system. The instrument measures the refractive index and thickness of transparent plates using a fibre-based low-coherence interferometer with a line-scan spectrometer. The spectrometer allows on-line dispersion measurement which is necessary to derive the sample thickness t as well as both the phase and group refractive indices n
p, n
g. The measurements were performed on a sample with six surfaces consisting of three glass windows mounted in a lens tube. Experimental results show that a measurement time of 4.4 s for the multi-layered object, which has a total thickness of approximately 10.5 mm, can be achieved whilst maintaining an accuracy of better than 0.1% for n
p, n
g, and t. This represents an approximately hundredfold improvement over previously published measurements.
Funder
Engineering and Physical Sciences Research Council
Subject
Applied Mathematics,Instrumentation,Engineering (miscellaneous)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献