Influence of scanbody design and intraoral scanner on the trueness of complete arch implant digital impressions: An in vitro study

Author:

Meneghetti Priscila CeolinORCID,Li Junying,Borella Paulo Sérgio,Mendonça GustavoORCID,Burnett Luiz HenriqueORCID

Abstract

This study aimed to compare the accuracy of full-arch digital implant impressions using seven different scanbodies and four intraoral scanners. A 3D-printed maxillary model with six implants and their respective multi-unit abutments was used for this study. Seven scanbodies (SB1, SB2, SB3, SB4, SB5, SB6, and SB7) and four intraoral scanners (Primescan®, Omnican®, Trios 3®, and Trios 4®) were assessed. Each combination group was scanned ten times and a dental lab scanner (D2000, 3Shape) was used as a reference. All scans were exported as STL files, imported into Convince software (3Shape) for alignment, and later into Blender software, where their 3D positions were analyzed using a Python script. The 3D deviation, angular deviation, and linear distance between implants #3 and #14 were also measured. Accuracy was measured in terms of “trueness” (scanbody 3D deviation between intraoral scan and desktop scan). Kruskal-Wallis followed by the Bonferroni correction was used to analyze the data (⍺ = .05). The study found statistically significant differences in digital impression accuracy among the scanners and scanbodies (p<0.001). When comparing different intraoral scanners, the Primescan system showed the smallest 3D deviation (median 110.59 μm) and differed statistically from the others, while Trios 4 (median 122.35 μm) and Trios 3 (median 130.62 μm) did not differ from each other (p = .284). No differences were found in the linear distance between implants #3 and #14 between Trios 4, Primescan, and Trios 3 systems. When comparing different scanbodies, the lowest median values for 3D deviation were obtained by SB2 (72.27μm) and SB7 (93.31μm), and they did not differ from each other (p = .116). The implant scanbody and intraoral scanner influenced the accuracy of digital impressions on completely edentulous arches.

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3