The Effect of High Nitric Oxide Concentrations on Oxygenators in Cardiopulmonary Bypass Machines (Experimental Study)

Author:

Radovskiy Alexey M.1,Vorotyntsev Ilya V.2,Atlaskin Artem A.2,Petukhov Anton N.2,Kryuchkov Sergey S.2,Atlaskina Maria E.2,Stepakova Anna N.2,Marichev Alexander O.1,Barygin Egor K.1,Osovskikh Victor V.1,Selemir Victor D.3,Buranov Sergey N.3,Golovanov Vladimir V.3,Shirshin Alexander S.3,Valueva Yulia V.3,Pichugin Vladimir V.4,Domnin Stepan E.5,Bautin Andrey E.1

Affiliation:

1. Almazov National Medical Research Center

2. Mendeleev University of Chemical Technology of Russia

3. Russian Federal Nuclear Center, All-Russian Scientific Research Institute of Experimental Physics

4. Privolzhsky Research Medical University

5. Research Institute “Specialized Cardiosurgical Clinical Hospital named after Academician B.A. Koroleva”

Abstract

The aim of the study. To study the effect of high nitric oxide concentrations on hollow polypropylene fibers of oxygenators.Materials and methods. The study was conducted in two stages. At the first stage, we evaluated the stability of oxygenator membrane made of hollow polypropylene fibers after six hours of exposure to air-oxygen mixture containing NO at 500 parts per million, or 500 pro pro mille (ppm) concentration, using mass spectrometry and infrared spectroscopy. At the second stage, an experiment with cardiopulmonary bypass (CPB) was conducted on 10 pigs. In the study group (n=5) animals sweep gas was supplied to the oxygenator as an air-oxygen mixture with NO at 100 ppm. In the control group animals (n=5) an air-oxygen mixture was used without NO. The CPB lasted for 4 hours, followed by observation for 12 hours. NO, NO2 (at the inlet and outlet of the oxygenator), and the dynamics of methemoglobin were evaluated. After weaning of animals from CPB, the oxygenators were tested for leakproofness, and scanning electron microscopy (SEM) was performed.Results. The oxygenator made of polypropylene hollow fibers retained its gas transfer parameters after six hours of exposure to air-oxygen mixture containing NO at 500 ppm. Based on IR-Fourier spectroscopy findings, NO did not affect structural integrity of polypropylene membranes. NO added to gas mixture at 100 ppm did not increase NO2 to toxic level of 2 ppm in 91% of control tests during 4 hours CPB in pigs; mean value was 1.58 ± 0.28 ppm. Methemoglobin concentration did not exceed the upper limit of permissible level (3%), and there were no statistically significant differences with the control group. All tested oxygenators have passed the leakproofness test. According to SEM findings, larger amounts of fibrin deposits were found in the control group oxygenators vs study group.Conclusion. There were no negative effects of NO at 500 ppm concentration on the oxygenator membrane made of hollow polypropylene fibers. NO at 100 ppm in a gas-mixture supplied to oxygenators did not lead to an exceedance of safe NO2 and methemoglobin concentrations in an animal model. Reduced fibrin deposits on hollow fibers of polypropylene oxygenator membranes were observed when with NO at a level of 100 ppm was added to a gas mixture.  

Publisher

FSBI SRIGR RAMS

Subject

Critical Care and Intensive Care Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3