Clinical implication by differential analytical performances of serum free light chain quantitation analysis using fully automated analyzers

Author:

Yun Shin Young1ORCID,Rim John Hoon1ORCID,Park Hak1,Kang Hyein1ORCID,Lee Sang-Guk1ORCID,Lim Jong-Baeck1ORCID

Affiliation:

1. Department of Laboratory Medicine , Yonsei University College of Medicine , Seoul , Korea

Abstract

Abstract Objectives Free light chain (FLC) is used for the diagnosis and prediction with regard to the progression risk of plasma cell disorders and Freelite reagent using the SPAplus analyzer (The Binding Site) has been one of the widely used option. However, N Latex FLC reagent with the Atellica CH 930 analyzer (Siemens Healthineers) has shown the advantages of automation and high throughput. We aimed to evaluated clinical implication by differential analytical performances of two assays. Methods A total of 322 serum samples were collected from 193 patients requested for FLC analysis including 131 multiple myeloma patients. The precision, linearity, dilution recovery of N Latex FLC assay was evaluated. We compared the two assays and analyzed the monomer-dimer pattern for discrepant results. Results The precision, linearity, and dilution recovery performance was appropriate for the routine use in clinical laboratories. Despite the good correlation within normal range, proportional bias up-to 170% was observed in samples with high concentrations especially for lambda. The higher value samples with N Latex FLC assay contained more monomer forms than controls. All opposite changes of FLC burden by the N Latex FLC assay proved to present concordant dynamic changes when assessed by serum protein electrophoresis. Conclusions Clinical laboratories should be aware of the inter-assay variability of FLC quantitative measurements using different platforms, especially for high concentrations of both kappa and lambda measurements, possibly due to monomer/dimer ratio diversity. Clinical interpretations for multiple myeloma disease status might not be dramatically affected only when the same assay is utilized during follow-up periods.

Funder

Siemens Healthineers

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3