Superb microvascular ultrasound is a promising non-invasive diagnostic tool to assess a ventriculoperitoneal shunt system function: a feasibility study

Author:

Brawanski Konstantin,Petr OndraORCID,Hernandez Christian Preuss,Kögl Nikolaus,Thomé Claudius,Gizewski Elke R.,Gruber Hannes,Verius Michael,Gruber Leonhard,Putzer Daniel

Abstract

AbstractThe objective of this pilot study was to assess the reliability of superb microvascular ultrasound (SMI) for the measurement of the cerebrospinal fluid (CSF) flow within VPS systems as an indirect sign for shunt dysfunction. Asymptomatic hydrocephalus patients, with a VPS system implanted between 2017 and 2021, were prospectively enrolled in the study. Using SMI, the CSF flow within the proximal and distal catheters were analysed. Before and after pumping the shunt reservoir, intraabdominal free fluid, optical nerve sheath diameter (ONSD), and papilla diameter (PD) were evaluated and correlated with the amount of valve activation. Nineteen patients were included. A flow was detectable in 100% (N = 19) patients in the proximal and in 89.5% (N = 17) in the distal catheter. The distal catheter tip was detectable in 27.7% (N = 5) patients. Free intraabdominal fluid was initially detected in 21.4% (N = 4) patients and in 57.9% (N = 11) at the end of the examination (P = 0.049). ONSD was significantly lower after pump activation (4.4 ± 0.9 mm versus 4.1 ± 0.8 mm, P = 0.049). Both peak velocity and flow volume per second were higher in proximal compared to distal catheters (32.2 ± 45.2 versus 5.6 ± 3.7 cm/sec, P = 0.015; 16.6 ± 9.5 ml/sec versus 5.1 ± 4.0 ml/sec, P = 0.001, respectively). No correlation was found between the number of pump activations and the changes in ONSD (P = 0.975) or PD (P = 0.820). SMI appears to be a very promising non-invasive diagnostic tool to assess CSF flow within the VPS systems and therefore affirm their function. Furthermore, appearance of free intraperitoneal fluid followed by repeated compression of a shunt reservoir indicates an intact functioning shunt system.

Funder

University of Innsbruck and Medical University of Innsbruck

Publisher

Springer Science and Business Media LLC

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3