Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract

Author:

Ma Shan,Gao Rongyu,Sun Jing,Yang Jun,Wen Kai,Chen Xiteng,Zhao Fangyu,Xu Xinyan,Tian FangORCID

Abstract

Abstract Purpose To assess the differences and similarities in the corneal curvature obtained by two swept-source optical coherence tomography (SS-OCT) devices, Scheimpflug imaging system and one ray tracing aberrometer in patients with cataracts. Moreover, this study aimed to compare the differences in posterior corneal (PK), total corneal (TK) and true net power (TNP) measurements among the IOLMaster 700, CASIA2, and Pentacam. Methods A total of 200 eyes of 200 patients (116 female, 58%) were enrolled in this study, with a mean age of 65.9 ± 9.5 years. The flattest (Kf), steepest (Ks), and mean cornal powers (Km), J0, and J45 were obtained using two SS-OCT-based biometric devices, one rotating camera system and one ray-tracing aberrometer. The PK, TK and TNP values were also measured using these devices. To evaluate the differences and similarities between the devicves, the Friedman test, Pearson correlation coefficient (r), intraclass coefficient correlation (ICC) and Bland‒Altman plots with 95% limits of agreement (LoA) were used, and boxplots and stacked histograms were generated to describe the distributions of the data. Results There were no significant differences between the IOLMaster 700 and Pentacam for any of the keratometry values. Additionally, there were no significant differences between the IOLMaster 700 and iTrace in evaluating J0 and J45. Bland‒Altman plots revealed relatively wide LoA widths, almost larger than 1 diopter for the keratometry values and almost larger than 0.5 diopter for J0 and J45 values among the four devices. In terms of PK and TK values, significant differences and low ICCs were found among the three devices. Conclusions Although strong correlations and good agreement were found among the IOLMaster700, CASIA2, Pentacam and iTrace for Kf, Ks, Km and J0, J45, it seems that the measurements should not be used interchangeably because of the wide LoA widths and the presence of significant differences among the devices. Similarly, due to significant differences and low ICCs, the PK, TK and TNP values obtained by IOLMaster 700, CASIA2, and Pentacam should not be used interchangeably.

Funder

Tianjin Key Medical Discipline (Specialty) Construction Project

Publisher

Springer Science and Business Media LLC

Subject

Cellular and Molecular Neuroscience,Sensory Systems,Ophthalmology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3