Particle therapy using protons or carbon ions for cancer patients with cardiac implantable electronic devices (CIED): a retrospective multi-institutional study

Author:

Hashimoto TakayukiORCID,Demizu Yusuke,Numajiri Haruko,Isobe Tomonori,Fukuda Shigekazu,Wakatsuki Masaru,Yamashita Haruo,Murayama Shigeyuki,Takamatsu Shigeyuki,Katoh Hiroyuki,Murata Kazutoshi,Kohno Ryosuke,Arimura Takeshi,Matsuura Taeko,Ito Yoichi M.,

Abstract

Abstract Purpose To evaluate the outcomes of particle therapy in cancer patients with cardiac implantable electronic devices (CIEDs). Materials and methods From April 2001 to March 2013, 19,585 patients were treated with proton beam therapy (PBT) or carbon ion therapy (CIT) at 8 institutions. Of these, 69 patients (0.4%, PBT 46, CIT 22, and PBT + CIT 1) with CIEDs (64 pacemakers, 4 implantable cardioverter defibrillators, and 1 with a cardiac resynchronization therapy defibrillator) were retrospectively reviewed. All the patients with CIEDs in this study were treated with the passive scattering type of particle beam therapy. Results Six (13%) of the 47 PBT patients, and none of the 23 CIT patients experienced CIED malfunctions (p = 0.105). Electrical resets (7) and over-sensing (3) occurred transiently in 6 patients. The distance between the edge of the irradiation field and the CIED was not associated with the incidence of malfunctions in 20 patients with lung cancer. A larger field size had a higher event rate but the test to evaluate trends as not statistically significant (p = 0.196). Conclusion Differences in the frequency of occurrence of device malfunctions for patients treated with PBT and patients treated with CIT did not reach statistical significance. The present study can be regarded as a benchmark study about the incidence of malfunctioning of CIED in passive scattering particle beam therapy and can be used as a reference for active scanning particle beam therapy.

Funder

Japan Society for the Promotion of Science

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Soft error measurement for carbon ion radiotherapy;Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms;2024-08

2. The Contribution of Secondary Particles Following Carbon Ion Radiotherapy to Soft Errors in CIEDs;IEEE Open Journal of Engineering in Medicine and Biology;2024

3. Pacemaker Malfunction During Passive Proton Beam Therapy for Localized Prostate Cancer: Case Reports and a Literature Review;Cureus;2023-09-29

4. Risk of cardiac implantable device malfunction in cancer patients receiving proton therapy: an overview;Frontiers in Oncology;2023-07-04

5. Neutron dose from a 6-MV X-ray beam in radiotherapy;Radiological Physics and Technology;2023-02-13