Evaluation of a System for Automatic Detection of Diabetic Retinopathy From Color Fundus Photographs in a Large Population of Patients With Diabetes

Author:

Abràmoff Michael D.123,Niemeijer Meindert34,Suttorp-Schulten Maria S.A.5,Viergever Max A.4,Russell Stephen R.12,van Ginneken Bram4

Affiliation:

1. Retina Service, Department of Ophthalmology and Visual Sciences, University of Iowa Hospitals and Clinics, Iowa City, Iowa

2. Department of Veterans Affairs, Iowa City VA Medical Center, Iowa City, Iowa

3. Department of Electrical and Computer Engineering, University of Iowa, Iowa City, Iowa

4. Image Sciences Institute, University Medical Center Utrecht, Utrecht, the Netherlands

5. Ophthalmology Service, OLVG Hospital, Amsterdam, the Netherlands

Abstract

OBJECTIVE—To evaluate the performance of a system for automated detection of diabetic retinopathy in digital retinal photographs, built from published algorithms, in a large, representative, screening population. RESEARCH DESIGN AND METHODS—We conducted a retrospective analysis of 10,000 consecutive patient visits, specifically exams (four retinal photographs, two left and two right) from 5,692 unique patients from the EyeCheck diabetic retinopathy screening project imaged with three types of cameras at 10 centers. Inclusion criteria included no previous diagnosis of diabetic retinopathy, no previous visit to ophthalmologist for dilated eye exam, and both eyes photographed. One of three retinal specialists evaluated each exam as unacceptable quality, no referable retinopathy, or referable retinopathy. We then selected exams with sufficient image quality and determined presence or absence of referable retinopathy. Outcome measures included area under the receiver operating characteristic curve (number needed to miss one case [NNM]) and type of false negative. RESULTS—Total area under the receiver operating characteristic curve was 0.84, and NNM was 80 at a sensitivity of 0.84 and a specificity of 0.64. At this point, 7,689 of 10,000 exams had sufficient image quality, 4,648 of 7,689 (60%) were true negatives, 59 of 7,689 (0.8%) were false negatives, 319 of 7,689 (4%) were true positives, and 2,581 of 7,689 (33%) were false positives. Twenty-seven percent of false negatives contained large hemorrhages and/or neovascularizations. CONCLUSIONS—Automated detection of diabetic retinopathy using published algorithms cannot yet be recommended for clinical practice. However, performance is such that evaluation on validated, publicly available datasets should be pursued. If algorithms can be improved, such a system may in the future lead to improved prevention of blindness and vision loss in patients with diabetes.

Publisher

American Diabetes Association

Subject

Advanced and Specialized Nursing,Endocrinology, Diabetes and Metabolism,Internal Medicine

Reference41 articles.

1. Fong DS, Aiello L, Gardner TW, King GL, Blankenship G, Cavallerano JD, Ferris FL II, Klein R: Diabetic retinopathy. Diabetes Care 26: 226–229, 2003

2. Centers for Disease Control and Prevention: Data & Trends: National Diabetes Surveillance System: Preventive care practices, 1994–2004. Available from http://www.cdc.gov/diabetes/statistics/ preventive/tX.htm. Accessed 8 July 2007

3. Fong DS, Aiello LP, Ferris FL III, Klein R: Diabetic retinopathy. Diabetes Care 27: 2540–2553, 2004

4. Fong DS, Aiello L, Gardner TW, King GL, Blankenship G, Cavallerano JD, Ferris FL III, Klein R: Retinopathy in diabetes. Diabetes Care 27(Suppl. 1): S84–S87, 2004

5. Chia DS, Yap EY: Comparison of the effectiveness of detecting diabetic eye disease: diabetic retinal photography versus ophthalmic consultation. Singapore Med J 45: 276–279, 2004

Cited by 198 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Diabetic retinopathy detection using ensembled transfer learning based thrice CNN with SVM classifier;Multimedia Tools and Applications;2024-01-30

2. Method of Diagnosis of Glaucoma Based on Human Fundus Images;Proceedings of the Southwest State University. Series: IT Management, Computer Science, Computer Engineering. Medical Equipment Engineering;2024-01-24

3. Teleophthalmology and retina: a review of current tools, pathways and services;International Journal of Retina and Vitreous;2023-12-05

4. Detection of Glaucoma and Diabetic Retinopathy Using Fundus Images and Deep Learning;2023 IEEE 5th International Conference on Cybernetics, Cognition and Machine Learning Applications (ICCCMLA);2023-10-07

5. Diabetic retinopathy detection using EADBSC and improved dilated ensemble CNN-based classification;Multimedia Tools and Applications;2023-09-21

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3